NASA says hammer-tap tests need not produce a half-sine pulse
A NASA technical bulletin says test quality depends on accurate force and response measurements and whether the excitation covers the test’s purpose.

Instrumented hammer-tap tests do not need to produce an ideal half-sine force waveform, according to NASA technical bulletin TB 26-09. The bulletin says a test should be assessed using the complete measured force and structural response.
The guidance applies to hammer tests used in modal analysis and tests that may precede pyroshock testing. NASA says changes in force shape affect the excitation but do not, by themselves, invalidate the record.
A half-sine pulse is broadband, but its spectrum has zeros at certain frequencies. Ripples in the force or separated hammer contacts can change the spectral content. NASA advises inspecting the measured spectrum alongside the response signal-to-noise ratio.
The bulletin recommends checking force measurement accuracy, bandwidth, calibration, dynamic range and time synchronisation. Response measurements should also have sufficient bandwidth and signal-to-noise ratio, without clipping or saturation.
Testers should confirm that the excitation covers the frequency, amplitude and nonlinear-response ranges required by the intended objective. NASA says a test should be rejected or repeated only when a specific measurement, excitation or coverage deficiency is identified.